1.gif (1892 bytes)

Brief Reports

Indian Pediatrics 2000;37: 771-775

Non-Invasive Estimation of Total Serum Bilirubin by Multi-Wavelength Spectral Reflectance in Neonates

Rakesh Lodha
Ashok K. Deorari
Vishal Jatana
Vinod K. Paul

From the Department of Pediatrics, All India Institute of Medical Sciences, New Delhi 110 029, India.

Reprint requests: Dr. Ashok K. Deorari, Additional Professor, Divison of Neonatology, Department of Pediatrics, All India Institute of Medical Sciences, New Delhi 110 029, India.

E-mail: [email protected]

Manuscript received: September 20, 1999;
Initial review completed: October 6, 1999;
Revision accepted: January 11, 2000

Neonatal jaundice occurs in 70% of term and 80% of preterm babies. Careful monitoring of all jaundiced newborns is required to ensure proper management. Total serum bilirubin (TSB) is commonly determined by spectro-photometric methods by analyzing plasma or serum sample. In addition to need for drawing a blood sample, there is wide range of vari-ability regarding the performance of bilirubin analyzers(1).

A large number of studies have demons-trated the possibility of prediction of serum bilirubin in neonates by analysis of the spectral reflectance from the skin(2–8). However, the accuracy of these techniques has been complicated by the variability introduced by skin pigmentation and the dermal maturity; so, the results of studies in white infants may not be applicable to heterogeneous Indian popula-tion. A hand-held transcutaneous bilirubino-meter (BiliCheckTM) has been designed to correct for these interfering factors; being based on recent studies on light scattering characteristics of the human skin(9). This instrument performs a spectral analysis by measuring the intensity of the reflected light at more than 100 individual wavelengths. The major skin components, which impact the spectral reflectance in newborns, are dermal maturity, melanin, hemoglobin and bilirubin(9). By subtracting the spectral contribution of the known components, the bilirubin absorbance is calculated. The present investigation was conducted to determine the clinical accuracy of this Multi-wavelength Spectral Reflectance (MWSR) device (BiliCheckTM) for measuring the transcutaneous bilirubin (TcB) in North Indian newborns.

Methods

This study was conducted at the Neonatal Unit of the Department of Pediatrics, AIIMS, New Delhi. Term neonates between ages of 2 and 7 days clinically having icterus more than 8 mg/dl were enrolled in the study. Newborns receiving phototherapy and those who under-went exchange transfusion were excluded. The postnatal age, gestation, birth weight and clinical level of icterus was recorded.

TcB measurements were taken on the forehead of each baby using MWSR device. Except for 38 babies, the tip of the instrument (BiliCal) was reused. Within 30 minutes of this estimation, a venous sample was drawn into a glass capillary tube, by another investigator unaware of the above value, for estimating the total serum bilirubin (TSB) value using twin beam Microbilimeter (Ginevri Technologie Biomediche). Both the instruments were cali-brated as recommended by the manufacturers.

The data was analyzed using software package Microsoft Excel. The values obtained by Microbilimeter were taken as the gold standard for calculating sensitivity and speci-ficity of transcutaneous bilirubin analyzer and clinical assessment of jaundice by pediatrician. Different tests of agreement were applied. The two sets of values were correlated. In addition the paired differences were calculated and the limits of agreement determined using the method suggested by Bland and Altman(10).

Results

One hundred and nine term neonates with non hemolytic jaundice were included; the mean (SD) gestation and birthweight were 38.4 (1.2) weeks and 2870 (330) grams, respec-tively. The estimations were done at mean (SD) age of 71.8 (27) hours.

A total of 121 paired estimations were performed. The TSB values ranged from 4.5 to 21.7 mg/dl. The TcB values correlated fairly with the TSB; the r being 0.823 (Fig. 1). In addition, the mean (SD) paired difference between the TSB and TcB was 0.01 (1.88) mg/dl; the range being –5.1 to 5.8 mg/dl. The 95% limits of agreement were estimated to be –3.6 to 3.6 mg/dl. According to these observations the two values agree poorly (Fig. 2).

The sensitivity of TcB to pick up TSB >13 mg/dl was 69%, with a specificity of 89.3% while, with eyes of experienced pediatrician sensitivity of 52.2% and specificity 89.3% was obtained (Table I ). The clinical assessment of icterus correlated better with TcB values (r = 0.84) than with the TSB (r = 0.67).

Forty-six babies had TSB greater than 13 mg/dl. In this subgroup the correlation between the TSB and TcB values was poor (r = 0.64). The mean (SD) paired difference between TSB and TcB was 1 (1.7) mg/dl; the 95% limits of agreement being –2.4 and 4.4 mg/dl.

We analyzed separately the data for 38 babies where fresh tips were used. The TcB values correlated fairly well with the TSB (r = 0.86). The mean paired difference between the TSB and TcB was 0.5 (1.6) mg/dl; the 95% limits of agreement being –2.7and 3.7 mg/dl. These results are not significantly different from the overall results. Also, the correlation was poor in these babies (n = 16) with bilirubin greater than 13 mg/dl (r = 0.69).

Fig. 1. Correlation between total serum bilirubin and transcutaneous bilirubin Fig. 2. Difference between total serum bilirubin and transcutaneous bilirubin.

              

Table I Comparison of Transcutaneous and Clinical Judgement of Bilirubin Levels with Total Serum Bilirubin

Total Serum bilirubin (mg/dl) Transcutaneous bilirubin analyser Clinical judgement
Sensitivity (%) Specificity (%) Sensitivity (%) Specificity (%)
> 13 69 89.3 52.2 89.3
  (55.7-82.3) (82.3-96.3 (37.8-66.6) (82.3-96.3)
> 15 46.6 9 53.3 98.1
  (26.7-66.6) (97-100) (33.3-73.3) (95.4-100)
³ 18 20 100 * *
  (0-40)      

* Clinical judgement can’t differentiate this group.

Figures in parentheses show the 95% confidence intervals.

Discussion

Often a pediatrician is interested to corroborate his clinical estimate of icterus with actual serum bilirubin level. Any technique to estimate the level of bilirubin in neonate should ideally be accurate, noninvasive and valid. The results from our study suggest that TcB estimations (BiliCheckTM) agree poorly with the TSB values in heterogeneous North Indian newborn population. Our study suggests that the instrument cannot replace TSB estimations. However, it may be useful as a screening test in centers where trained pediatrician is not available as the TcB correlated fairly well with the clinical estimation.

Bhutani et al. reported a high degree of correlation between TcB values (BiliCheckTM) and TSB in American babies(11). At present no other studies are available to confirm these findings. The other transcutaneous bilirubino-meters measure the yellowness of the neonate’s skin, making no allowance for skin maturity or melanin. While these have been shown to be useful in white and certain other populations(2-8), the Indian studies have not been able to show the same degree of accuracy(12–14). Elsewhere also poor correlation has been reported(15). Even the laboratory analyzers have an undesirable intra- and inter-instrument variability(1).

We also found that the agreement between TSB and TcB was poor in the subgroup where the TSB was more than 13 mg/dl. Similar results were reported earlier using a different transcutaneous bilimeter(16). The sensitivity of clinical judgement for values >13 mg/dl was poor, though the specificity was good. It seems that the TcB estimation reconfirms the clinical judgement specially for higher bilirubin levels in pathological range (>13 mg/dl).

The present study also demonstrated the drawbacks of using correlation as the sole measure of agreement. While the correlation may be good, the difference between the two values may be too large to be acceptable clinically. This has been well highlighted by Bland and Altman(10). In none of the earlier studies, the differences between the two sets of values were looked at.

While the manufacturer recommends that the tip (BiliCalTM) be used only once, we reused the tip in majority of the babies. However, the subgroup analysis did not find significant difference between single use and multiple use. We conclude that this TcB instrument may be useful as a screening tool, but it cannot substitute for TSB estimations.

Acknowledgement

We thank Medisphere Marketing Ltd; New Delhi for providing us the instrument and accessories to conduct the study.

Contributors: AKD, VKP coordinated the study (design and interpretation); RL and VJ helped in data collection and RL drafted the paper. AKD will act as the guarantor for the study.

Funding: None.

Competing: None.

Key Messages


Otogenic route is the commonest mode of acquiring post-neonatal tetanus.

• Otogenic tetanus occurs mainly below 6 years while post-injury tetanus commonly occurs above 6 years of age.

• Lower respiratory tract infection and autonomic imbalances are common complications in tetanus.

• Mortality in tetanus is related to short period of onset, long duration of spasms, increasing severity of tetanus and presence of complications.

• Tetanus immunization should always be checked in cases with otitis media.

 

References

1. Vreman HJ, Verter J, Oh W, Fanaroff AA, Wright LL, Lemons JA, et al. Interlaboratory variability of bilirubin measurements. Clin Chem 1996; 42: 869-873.

2. Yamanouchi I, Yamauchi Y, Igarshi I. Transcutaneous bilirubinometry: Preliminary studies of non invasive transcutaneous bilirubin meter in the Okayama national hospital. Pediatrics 1980; 65: 192-202.

3. Hanneman RE, Schreiner RL, Dewitt DP, Norris SA, Glick MR. Evaluation of the Minolta bilirubin meter as a screening device in white and black infants. Pediatrics 1982; 69: 107-109.

4. Tan KL. Transcutaneous bilirubinometry in full term Chinese and Malay infants. Acta Paediatr Scand 1982; 71: 593-596.

5. Yamauchi Y, Yamanouchi I. Transcutaneous bilirubinometry: Evaluation of accuracy and reliability in a large population. Acta Paediatr Scand 1988; 77: 791-795.

6. Knudsen A. Measurement of the yellow color of the skin as a test of hyperbilirubinemia in mature newborns. Acta Paediatr Scand 1990; 79: 1175-1181.

7. Tan KL, Chia HP, Koh BC. Transcutaneous bilirubinometry in Chinese, Malay and Indian infants. Acta Paediatr Scand 1996; 85: 986-990.

8. Tayaba R, Gribetz D, Gribetz I, Holzman IR. Noninvasive estimation of serum bilirubin. Pediatrics 1998; 102: E28.

9. Jacques S, Saidi I, Ladner A, Oelberg D. Developing an optical fiber reflectance spectrometer to monitor bilirubinemia in neonates. SPIE Proceedings 2975, Laser-Tissue Interactions, San Jose, CA February 1997; pp 1-7.

10. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; i: 307-310.

11. Bhutani V, Johnson L, Gourley G, Dwaranczyk R, Grous M. Non invasive measurement of total serum bilirubin by multiwavelength specteral reflectance by BiliCheckTM in newborn patients. Pediatr Res 1998; 43: 167A.

12. Kumar A, Faridi MM, Singh N, Ahmad SH. Transcutaneous bilirubinometry in the management of bilirubinemia in term neonates. Indian J Med Res 1994; 99: 227-230.

13. Harish R, Sharma DB. Transcutaneous bilirubinometry in neonates: Evaluation of Minolta Air Shields jaundice meter. Indian Pediatr 1998; 35: 264-267.

14. Bhat V, Srinivasan S, Usha TS, Puri RK. Correlation of transcutaneous bilirubinometry with serum bilirubin in South Indian neonates. Indian J Med Res 1987; 86: 49-52.

15. Bourchier D, Cull AB, Oettli PE. Transcutaneous bilirubinometry: 22 months experience at Waikato Women’s Hospital. N Z Med J 1987; 100: 599-600.

16. Taha SA, Karrar ZA, Dost SM. Transcutaneous bilirubin measurement in evaluating neonatal jaundice among Saudi newborns. Ann Trop Paediatr 1984; 4: 229-231.

Home

Past Issue

About IP

About IAP

Feedback

Links

 Author Info.

  Subscription