1.gif (1892 bytes)

Personal Practice

Indian Pediatrics 1998; 35:631-640

Recent Advances in Assisted Ventilation for Neonatal Respira- tory Distress Syndrome


Victor Y.H. Yu

Reprint requests: Dr. Victor Y.H. Yu, Professor of Neonatology, Monash University, Director of Neonatal Intensive Care, Monash Medical Center, 246, Clayton Road, Clayton, Victoria 3168, Australia.

This manuscript provides an overview of recent advances in assisted ventilation in the neonate including continuous distending pressure (CDP), intermittent mandatory ventilation (IMV), synchronised ventilation, sedation and muscle relaxation, weaning and extubation, and high frequency ventilation. Although a great deal of research has been carried out on the treatment of respiratory distress syndrome (RDS), there is no 'best way' of clinical management as there remain many areas of uncertainty and often little objective evidence to indicate that one approach is better than another(1).

Continuous Distending Pressure

Techniques used to generate continuous positive airway pressure (CP AP) include a headbox with neck seal, face-mask, short twin nasal cannulae, long nasopharyngeal tube and endotracheal tube. Oropharyngeal pressures are similar whether monoor binasal cannulae are used(2). However, two short wide prongs should provide better pressure delivery than one long prong, as resistance is proportional to length and inversely proportion to radius. The Infant Flow Nasal CP AP System (flow driver), designed to assist expiratory flow, decrease respiratory resistance and decrease the work of breathing, is currently commonly used for CDP delivery. It must be applied correctly as nasal deformities including snubbing of the nose, flaring of the nostrils, and columella and septal necrosis have been described as complications from its use(3). A body-box which generates continuous negative extrathoracic pressure was used in the rnid-1970s but has recently been reintroduced for use in conjunction with IMV(4).

The benefits of CDP used for the treatment of RDS have been evaluated by meta-analysis(5). COP significantly improves oxygenation, reduces the need for subsequent IMV and lowers the mortality rate. However, the incidence of air leak is significantly increased and there is no reduction in chronic lung disease (CLD). The same meta-analysis show that, comparing early COP (when the FiO2 has reached 0.3- 0.6) to late COP (when the FiO2 has exceeded 0.6), early CDP results in a significant reduction in the need for subsequent IMV and duration of assisted ventilation, even though the reduction in the incidence of air leak and mortality did not reach statistical significance. A retrospective study comparing the incidence of CLD using different respiratory management policies and observational studies without a control group also suggested that the early use of nasal CDP in infants with RDS reduces the CLD rate(6). Although an observational study suggested that early nasal COP may be as good as initial endotracheal intubation and mechanical ventilation(7), a randomized controlled trial (RCT) is required to answer the question of which treatment is the best.

Intermittent Mandatory Ventilation

Pressure-limited ventilators remain by far the most commonly used in neonatal ventilation, and are also simpler to use and less expensive. However, with the current theory that ventilator induced lung injury does in fact result from 'volutrauma', there is renewed interest in volume-controlled ventilation using new ventilators designed with improved microprocessor technology(8).

Recent studies have demonstrated that if the peak inspiratory pressure (PIP) is held constant, this may decrease tidal volume and increase PaC02(9) as it reduces the difference between PIP and positive end-expiratory pressure (PEEP), and reduces lung compliance. PEEP is four times as potent as PIP in bringing about changes in tidal volume(1O). For example, reducing PEEP by 1 cm HP is twice as potent as increasing PIP by 2 cm Hp in achieving an increase in tidal volume. Similarly, a 1 cm H2O increase in PEEP is twice as effective as a 2 cm H20 decrease in PIP in reducing tidal volume. Excessive PEEP also over inflates the lung which increases pulmonary vascular resistance and intra-pulmonary right-to-Ieft shunting as well as impairs venous return and cardiac output(11)
.

Controversy remains whether a slow or fast ventilator rate is better, and the optimal inspiratory time (IT), expiratory time (ET) and inspiratory: expiratory ratio (IE ratio). Recently, a large multicenter trial was published involving over 300 infants(12), in which slow rate (initial settings 30/min, IE ratio 1:1, IT 1 sec, ET 1 sec) was compared to fast rate (initial settings 60/ min, IE ratio 1:2, IT 0.33 see, ET 0.67 sec). With the fast rate regime, when the mean airway pressure (MAP) reached 12 cm H20 and the FiO2 exceeded 0.6, the IE ratio was increased to 1:1 (IT 0.5 see, ET 0.5 sec) and the gas flow was increased from 6-8 1/ min to 8-10 1/ min. PIP and PEEP were similar in the two groups. The meta-analysis of all slow versus fast rate RCTs(5) showed a significant increase in the incidence in pneumothorax (PTX) with the slow rate regime for which the trend towards a lower incidence of CLD was offset by a trend towards a higher mortality rate. It appears that increasing the IT beyond 0.5 second on the slow rate regime may improve oxygenation by increasing MAP but the risk of PTX is also increased. However, infants with severe RDS cannot be adequately oxygenated with a short IT without resorting to a PIP of >30 cmH2O. Neither the fast or slow rate regime suits all infants, probably due to the fact that it depends on the type and severity of their underlying lung disease. With the fast rate regime, rates of 120/min have been shown to be safe in the majority of non-paralysed infants but this should be avoided in paralysed infants especially those >31 weeks gestation because it causes air trapping(13).

Synchronized Ventilation

Neonatal microprocessor controlled ventilators have been designed which allow the delivery of mechanical breaths in response to a signal derived from the infant's spontaneous inspiratory effort (synchronised ventilation). The trigger signal can be flow or pressure (measured within the ventilator circuit nearest the infant's airway) or impedance (thoracic or abdominal). Successful triggering depends on two factors: magnitude (the amount of infant effort required to trigger) and delay (the time from initiation of infant effort and the rise in proximal airway pressure). Flow triggering is the best(14) and transthoracic impedence is worse of the available techniques(15). The modes of synchronised ventilation include the following:

Assist/Control Ventilation

Also known as patient triggered ventilation (PTV), this combines a ventilator delivered positive pressure breath in response to the infant's inspiratory effort (assist) with a guaranteed mechanical breath at a preset rate if no infant effort is detected (control). Earlier clinical experience has not shown it to be useful in infants < 28 weeks gestation, in the first 24 hours after birth, in those requiring a high Fi02 and ventilator rate, and in ventilator dependent infants with CLD(16). Its role might have been in
I the larger, more mature infant with mild RDS during the recovery period especially when there is asynchronous breathing. One RCT has found that PTV is less successful in promoting synchrony than a fast rate regime on conventional ventilation(17). However, a RCT showed that infants of 1100-1500g with RDS on PTV utilizing a flow-derived signal to initiate and terminate a mechanical breath, had more rapid weaning that those on conventional pressure limited time-cycled ventilation(18). One study reported that PTV utilizing an airway pressure sensor is capable of providing long-term respiratory support from birth to extubation in even extremely preterm infants(19).

Synchronized Intermittent Mandatory Ventilation

In the SIMV mode, the mechanically delivered breaths are preset at a fixed rate but are synchronized to the onset of the infant's own breaths. For example, if the SIMV is 20 bpm and the infant is breathing at 60/min, the maximum number of triggered. breaths is 20. The synchrony produced by SIMV has been shown to result in a lower respiratory rate and a higher Pa02, and deliver larger and more consistent tidal volumes than during IMV(20-23). One RCT demonstrated benefits of SIMV over conventional IMV: a shorter duration of ventilation in infants> 2000 g, improved oxygenation in infants 1000-2000 g, and a lower CLD rate defined as supp.emental oxygen at 36 weeks postconceptional age in infants < 1000 g(24). The RCT comparing PTV and SIMV did not show any difference in wean- ing by pressure reduction (PTV) or rate reduction (SIMV)(25). Neither has it been shown that weaning by a combination of pressure and rate reduction, as can be achieved during SIMV, offer any advantage over PTV(26).

Pressure Support Ventilation (PSV)

PSV has been defined as infant initiated, pressure targeted and infant controlled ventilation, designed to assist the infant's spontaneous breathing with an inspiratory pressure 'boost'. Once the breath is triggered by the infant's inspiratory effort, a preset pressure is achieved and maintained throughout inspiration until the inspiratory flow falls below a preset value. Although, very similar to PTV, the infant has control of the inspiratory time. When combined with SIMV, PSV reduces the work of breathing which might be beneficial in weaning especially for infants with CLD(27,28). As new neonatal ventilators become available with advanced micro- processor based technology, ventilatory regimes based PTV, SIMV and PSV might prove one day to be superior to conventional IMV(29). Currently, RCTs comparing the newer ventilatory modes remain limited.

Sedation and Muscle Reiaxation

Muscle relaxants are often used in un- stable infants with RDS while on the ventilator. The concern is that spontaneous respiratory efforts which are out of phase with the ventilator breaths (active expiration) may increase the risk of PTX and intracranial hemorrhage. Sedation can be achieved with a narcotic such as morphine (100
µg/ kg q 4-6 h IM/IV, or 100 µg/kg/h over 2 hours followed by 25 µg/kg/h IV infusion), phenobarbitone, fentanyl or midazolam.(30-31). Morphine has been shown to encourage synchrony of the infant's breathing with the ventilator(32) but one RCT showed that ventilated infants given phenobarbitone had a higher incidence of air leak(33). Morphine results in small fall in heart and respiratory rates(34), but it has no significant effect on arterial blood pressure and cardiac and cerebral hemodynamics(35). Diamorphine is more potent than morphine and a loading dose of 50 µg/kg followed by an infusion of 15 µg/kg/h have been recommended in ventilated infants(36). Fentanyl (5 mg/kg loading and 1-2 µg/kg/h infusion)(37) and alfentanil (15-20 µg/kg loading and 3-5 µg/kg/h infusion)(38) are synthetic opioids which are even more potent and have the theoretical advantage of a shorter duration of action and greater cardiovascular stability. An alternative to sedation is to use fast ventilator rates (60-120/min) to achieve greater synchrony and less active expiration, or to use an IT and ET on the ventilator which matches those observed during spontaneous breating over a brief period on CDP(39). Preterm infants are capable of hormonal stress responses which might be detrimental to their outcome(40). Morphine or fentanyl infusion has been shown to reduce stress markers in preterm infants ventilated for RDS, and can be justified on humanitarian grounds for reducing the level of their pain and stress(37,40-42).

Neuromuscular paralysis is achieved with pancuronium (0.1 mg/kg q 3-4 h pm). The evidence that muscle paralysis prevents PTX, intraventricular haemorrhage (IVH) and CLD remains poor. Pancuronium should be used sparingly, one indication being infants> 34 weeks on mechanical ventilation who make vigorous respira
tory efforts which are detrimental to gas exchange(43). The risks include an increase in heart rate and blood pressure, deterioration of Pa02 in 20% of cases, progressive worsening of lung compliance and resistance, fluid retention and disuse muscle atrophy. The deterioration after pancuronium in some infants might be a result of a reduction of minute ventilation following cessation of spontaneous respirations, but it has been shown that the decrease in oxygenation is caused by alveolar derecruitment and reduction of FRC from loss of expiratory braking nJchanisms(44). Studies which compared pancuronium with morphine or pethidine reported that neither paralysis nor sedation had an adverse effect on heart rate or blood pressure and showed no difference in the duration of ventilation and the incidence of air leak, IVH or CLD(41-45). One RCT indicated that routine paralysis offers no advantage to fast rate ventilation which is the preferred strategy for ventilating infants with RDS(46).

Weaning and Extubation

RCTs showed that PTV or SIMV can re- duce the duration of weaning compared with conventional IMV(18,24-47). Prophylactic aminophylline therapy (10 mg/kg IV followed by 2.5 mg/kg q 12h) has been shown to increase the success rate of extubation, decrease the re-intubation rate, or prevent post-extubation apnea(48). Caffeine is as efficacious as aminophylline, and has the advantages of a wider therapeutic to toxic ratio, the need for only once a day administration, and less problem with tachycardia and feed intolerance(49). Doxapram may aid in weaning when given with amonophylline in infants who are unresponsive to aminophylline alone(50).

Post-extubation nasal CP AP has been shown in two recent RCTs to result
in a higher success rate of earlier extubation, lower respiratory rate, better blood gases and pH, less atelectasis and a lower re-intubation rate(51,52) but another three RCTs showed no benefit compared to extubation directly into a headbox(53-55), suggesting that the decision has been shown to reduce the problem of apnea of prematurity(56). There is no RCT evidence that dexamethasone is effective in preventing post-extubation laryngeal edema(57). However, one RCT reported earlier extubation in infants with severe RDS given dexamethasone therapy for the first 12 days after birth., starting on a dose of 1 mg/kg/ d(58). Studies which conducted pulmonary function testing to help determine the optimal timing of extubation in infants with RDS gave conflicting results as to its usefulness(59-62).

High Frequency Ventilation

Conventional mechanical ventilation uses rates up to 120/min and tidal volumes of 5 ml/kg, delivered by a ventilator with a pneumatic valve with a passive expiratory phase. High frequency jet ventilation (HFJV) uses rates of about 400/min and tidal volumes of about 3 ml/kg, delivered as compressed gas via a jet cannula with also a passive expiratory phase. High frequency oscillation (HFO) uses rates of 10- 15Hz (600-900/min) and tidal volumes of < 3ml/kg, delivered by a piston or loud- speaker to provide a continuous bias flow, and has an active expiratory phase. High frequency flow interrupted ventilation (HFFIV) mimics HFO but does not have an active expiratory phase. During high frequency ventilation, swings in alveolar pressure are extremely swing and pulmonary barotrauma is potentially reduced. Necrotising tracheobronchitis and gas trapping have been reported with HFJV but not With HFO. HFO can however result in a 14-18% reduction in left ventricular output(63). Compared to conventional ventilation, RCTs have not shown better(64), similar(65) and worse(66) outcomes with HFJV, and similar. outcome with HFFIV(67).

Six RCTs have been conducted with HFO and published in full: the first showed no benefits and a higher incidence of air leak, IVH and periventricular leukomalacia (PVL) with HFO(68); the second showed no difference in incidence of CLD but a less severe degree of pulmonary dysfunction among the CLD infants(69); the third in infants < 1750g with RDS showed a significant reduction in the incidence of CLD with no increased risk of IVH(70); the fourth in infants 750-2000g showed no difference in the rates of air leak, IVH, PVL, CLD or mortality(71); the fifth in infants with severe RDS showed a significant reduction in the incidence of air leak(72) and the sixth in infants < 36 weeks with RDS showed a number of benefits which include better oxygenation, less surfactant use, less CLD and NEC, and lower hospital costs(73), The first RCT was criticised for its low volume/ pressure strategy to HFO(74) as centers which use a high volume/pressure approach to HFO are reporting a reduction in air leak and CLD without the serious side effects such as IVH or PVL which was attributed to HFO(70,72,73). The MAP during HFO is known to be an important determinant of oxygenation(75) and earlier animal experiments have also indicated that maintenance of an adequate lung volume with a high MAP during HFO is essential to prevent atelectasis and lung injury. It was recommended that the MAP at the initiation of HFO be set at 2-3 cmH20 higher than that delivered by conventional mechanical ventilation. This MAP is measured at the junction of the ventilator circuit and the endotracheal tube, which is therefore an overestimate of mean alveolar pressure(76). Infants with atelectatic lungs who have the lowest lung volume require the greatest increase in MAP on transfer to HFO to optimise oxygenation(77).

HFJV has been shown in a RCT for treatment of pulmonary interstitial emphysema (PIE) to improve blood-gases, MAP, duration of PIE and mortality(78). An observational study also showed benefits of HFO in the rescue of infants with PIE(79). Reports also suggested benefits in using HFO for neonatal lobar emphysema, pulmonary hemorrhage and infants with respiratory failure secondary to increased intra-abdominal pressure(80-82). However, HFO is more effective in RDS or persistent pulmonary hypertension than in pulmonary airleak or pulmonary hypoplasia(83- 84). HFO has an important role as rescue
. treatment in severe neonatal lung disease when conventional ventilation fails. How- ever, its routine use as a primary mode of ventilation is a contentious issue until further clinical trials are performed(85).
 

 References


1. Joint Working Group of the British Association of Perinatal Medicine and the Research Unit of the Royal College of Physicians. Development of audit measures and guidelines for good practice in the management of neonatal respiratory distress syndrome. Arch Dis Child 1992; 67: 1221-1227.

2. Pedersen JE, Nielsen K. Oropharyngeal and esophageal pressures during mono and binasal CP AP in neonates. Acta Paediatr 1994;'83: 143-149.

3.
Robertson NJ, McCarthy LS, Hamilton P A, Moss ALH. Nasal deformities resulting from flow driver continuous positive airway pressure, Arch Dis Child 1996; 75: F209-F212.

4. Cvetnic WG, Cunnigham MD, Sills JH, Gluck L. Reintroduction of continuous negative pressure ventilation in neonates: Two-year experience. Pediatr Pulmonol 1990; 8: 245-253.

5. Bancalari E, Sinclari Je. Mechanical ventilation. In: Effective Care of the Newborn Infant. Ed. Sinclair JC, Bracken MB Oxford, Oxford University Press, 1992; pp 200-220.

6. Kamper J, Wuff K, Larsen e. Lindequist S. Early treatment with nasal continuous positive airway pressure in very low birth weight infants. Acta Paediatr 1993; 82: 193-197.

7.
Johnson B, Katz-Salamon M, Faxelius G, Broberger U, Lagercrantz H. Neonatal care of very low birthweight infantsin special care and neonatal intensive care units in Stockholm. Early nasal continuous positive airway pressure versus mechanical ventilation: Gains and losses. Acta Pediatr 1997; 86: 4-10.

8. Nicks
JJ, Becker MA, Donn SM. Ventilatory management casebook. J Perinatal 1993; 13: 72-75.

9. Greenough A. Chan V, Hird MF. Positive and expiratory pressure in acute and chronic respiratory distress. Arch Dis Child 1992; 67: 320-323.

10. Bartholomew KM, Brownlee KG, Snowden S, Dear PRF. To PEEP or not to PEEP? Arch Dis Child 1994; 70: F209- F212.

11. Evans N, Kluckow M. Early determinants of right and left ventricular output in ventilated preterm infants. Arch Dis Child . 1996;74:F88-F94.

12. Oxford Region Controlled Trial of Artificial Ventilation (OCTAVE). Multi-center randomized controlled trial of high versus low frequency positive pressure ventilation in 346 newborn infants. Arch Dis Child 1991; 66: 770-775.

13. Hird M, Greenough A, Gamsu H. Gas trapping during high frequency positive pressure ventilation using conventional ventilators. Early Hum Dev 1990; 22-: 51- 56.

14. Uchiyama A, Imanaka H, Taenaka N, Nakano S, Fujino Y, Yoshiya I. A comparative evaluation of pressure-triggering and flow-triggering in pressure support ventilation for neonates using an animal model. Anaes Intens Care 1995; 23: 302- 306.

15. Nikischin W, Gerhardt T, Everett R, Gonzalez A, Hummler H, Bancalari E. Patient-trtggered ventilation: A comparison of tidal volume and chest wall and abdominal motion as trigger signal. Pediatr Pulmonol1996; 22: 28-34.

16. Greenough A, Milner AD. Respiratory support using patient triggered ventilation in the neonatal period. Arch Dis Child 1992; 67: 69-71.

17. Hird MF, Greenough A. Randomized trial of patient triggered ventilation versus high frequency positive pressure ventilation in acute respiratory distress.
J Perinat Med 1991; 19: 379-384.

18. Donn SM, Nicks JJ, Becker MA. Flow-synchronized ventilation of preterm infants with respiratory distress syndrome.
J Perinatol1994; 14: 90-94.

19. de Boer RC, Jones A, Ward PS, Baumer JH. Long term trigger ventilation in neo- natal respiratory distress syndrome. Arch Dis Child 1993; 68: 308-311.

20. Bernstein G, Heldt GP, Mannino FL. Increased and more consistent tidal volumes during synchronized intermittent mandatory ventilation in newborn infants. Am
J Respir Crit Care Med 1994; 150; 1444-1448.

21. Mizuno K, Takeuchi T, Itabashi K, Okuyama K. Efficacy of synchornized IMV on weaning neonates from the ventilator. Acta Paediatr Jap 1994; 36: 162-166.

22. Cleary JP, Bernstein G, Mannino FL, Heldt GP. Improved oxygenation during synchronized intermittent mandatory ventilation in neonates with respiratory distress syndrome: A randomized crossover study.
J Pediatr 1995; 126: 407-411.

23. Smith KM, Wahlig TM, Bing DR. Georgieff MK, Boros SJ, Mammel Me. Lower respiratory rates without decrease in oxygen consumption during neonatal synchronized intermittent mandatory ventilation. Intens Care Med 1997; 23: 463-438.

24. Bernstein G, Mannino FL, Heldt GP, Callahan JD, Bull DH, Sola A, et al. Randomized multicenter trial comparing synchronized and conventional intermittent mandatory ventilation in neonates.
J Pediatr 1996; 128: 4S3-463.

25. Chan V, Greenough A. Comparison of weaning by patient triggered ventilation or synchronous intermittent ventilation in preterm infants. Acta Paediatr 1994; 83: 335-337.

26. Dimitriou G, Greenough A, Griffin F, Chan V. Synchronous intermittent mandatory ventilation modes compared with patient triggered ventilation during weaning. Arch Dis Child 1995; 72: F188- F190.

27. EI-Khatib MF, Chatburn RL, Potts DL, Blumer JL, Smith PG. Mechanical ventilators optimized for paediatric use decrease work of breathing and oxygen consumption during pressure-support ventilation. Crit Care Med 1994; 22: 1942-1948.

28. Nicks
H, Becker MA, Donn SM. Bronchopulmonary dysplasia: Response to pressure support ventilation. J Perinatol1994; 14: 495-497.

29. Sinha SK, Donn SM. Advances in neonatal conventional ventilation. Arch Dis Child 1996; 75: F135-F140.

30. Jacqz-Aigrain E, Daoud P, Burtin P, Desplanques L, Beaufils F. Placebo-con- trolled trial of midazolam sedation in mechanically ventilated newborn babies. Lancet 1994; 344: 646-650.

31. Harte GJ, Gray PH, Lee TC, Steer P A, Charles BG. Hemodynamic responses and population pharmacokinetics of midazolam following administration to ventilated, preterm neonates.
J Pediatr Child Health 1997; 33: 335-338.

32. Dyke MP, Kohan R, Evans S. Morphine increases synchronous ventilation in preterm infants. J Pediatr Child Health 1995; 31: 176-179.

33. Elias-Jones AC, Barrett DA, Rutter N, Shaw PN, Davis SS. Diamorphine infusion in the preterm neonate. Arch Dis Child 1991; 66: 1155-1157.

34. Dyke MP, Kohan R, Evans S. Morphine increases synchronous ventilation in preterm infants.
J Pediatr Child Health 1995; 31: 176-179.

35. Sabatino G, Quartuli L, Di Fabio S, Ramenghi LA. Hemodynamic effects of intravenous morphine infusion in ventilated preterm babies. Early Hum Dev 1997; 263-270.

36. Barker DP, Simpson J, Pawula M, Barrett DA. Shaw PN, Rutter N. Randomized, double blind trial of two loading dose regimens of diamorphine in ventilated newborn infants. Arch Dis Child 1995; 73: F22-F26.

37. Orsini AJ, Leef KH, Costarino A, Dettorre MD, Stefano JL. Routine use of fentanyl infusions for pain and stress reduction in infants with respiratory distress syndrome.
J Pediatr 1996; 129: 140-145.

38. Marlow N, Weindling AM, Van Peer A, HeyKants
J. Alfentanil pharmacokinetics in preterm infants. Arch Dis Child 1990; 65: 349-351.

39. Amitay M, Etches PC, Finer NN, Maidens JM. Synchronous mechanical ventilation of the neonate with respiratory disease. Crit Care Med 1993; 21: 118-124.

40. Barker DP, Rutter N. Stress, severity of illness, and outcome in. ventilated preterm infants. Arch Dis Child 1996; 75: F187- F190.

41. Quinn MW, Otoo F, Rushforth JA, Dean HG, Puntis JW, Wild J, et al. Effect of morphine and pancuronium on the stress response in ventilated preterm infants. Early Hum Dev 1992; 30: 241-248.

42. Ionides SP, Weiss MG, Angelopoulos M, Myers IF, Handa RJ. Plasma beta-endorphin concentrations and analgesia-muscle relaxation in the newborn infant supported by mechanical ventilation.
J Pediatr 1994; 125: 113-116.

43. Levene MI, Quinn MW. Use of sedatives and muscle relaxants in newborn babies receiving mechanical ventilation. Arch Dis Child 1992; 67: 870-873.

44. Miller J, Law AB, Parker RA, Sundell H, Silberberg AR, Cotton RB. Effects of morphine and pancuronium on lung volume and oxygenation in premature infants with hyaline membrane disease.
J Pediatr 1994; 125: 97-103.

45. Shaw NJ, Cooke RW, Gill AB, Shaw NJ, Saeed M. Randomized trial of routine versus selective paralysis during ventilation for neonatal respiratory distress syndrome. Arch Dis Child 1993; 69: 479-482.

46. Quinn MW, Wild J, Dean HG, Hartley R, Rushforth JA, Puntis JW, et al. Randomized double-blind controlled trial of effect of morphine on catecholamine concentrations in ventilated preterm babies. Lancet 1993; 342: 324-327.

47. Chan V, Greenough A. Randomized controlled trial of weaning by patient trigerred ventilation or conventional ventilation. Eur
J Pediatr 1993; 152: 15-54.

48. Barrington KJ, Finer NN. A randomized controlled trial of aminophylline in ventilatory weaning of premature infants. Crit Care Med 1993; 21: 846-850.

49. Larsen PB, Brendstrup L, Skov L, Flachs H. Aminophylline versus caffeine citrate for apnea and bradycardia prophylaxis in premature neonates. Acta Paediatr 1995; 84: 360-364.

50. Brion LP, Vega-Rich C, Reinersman G, Roth P. Low-dose doxapram for apnea unresponsive to aminophylline in very low birthweight infants.
J Perinatol1991; 11: 359-364.

51. Higgins RD, Richter SE, Davis JM. Nasal continuous positive airway pressure
facilitates extubation of very low birth weight neonates. Pediatrics 1991; 88: 999- 1003.

52. So B-H, Tamura M, Mishina J, Watanabe T, Kamoshita S. Application of nasal con- tinuous positive airway pressure' to early extubation in very low birthweight infants. Arch Dis Child 1995; 72: FI91-FI93.

53. Chan V, Greenough A. Randomized trial of methods of extubation in acute and chronic respiratory distress. Arch Dis Child 1993; 68: 570-572.

54. Annibale DJ, Hulsey TC, Engstrom PC, Wallin LA, Ohning BL. Randomized controlled trial of nasopharyngeal continuous positive airway pressure in the extubation of very low birth weight infants.
J Pediatr 1994; 124: 455-460.

55. Tapia JL, Bancalari A, Gonzalez A, Mercado Me. Does continuous positive airway pressure during weaning from intermittent mandatory ventilation in very low birth weight infants have risks or benefits? A controlled trial. Pediatr Pulmonoll995; 19: 269-274.

56. Kurlak La, Ruggins NR, Stephenson TJ. Effect of nursing position on incidence, type, and duration of clinically significant apnea in preterm infants. Arch Dis Child 1994; 71: FI6-FI9.

57. Tellez DW, Galvis AG. Storgion SA, Amer HN, Hoseyni
. M, Deakers TW. Dexamethasone in the prevention of postextubation stridor in children. J Pediatr 1991; 118: 289-294.

58. Yeh TF, Torre JA, Rastogi A, Anyebuno MA, Pildes RS. Early postnatal dexamethasone therapy in premature infants with severe respiratory distress syndrome: A double-blind, controlled study. J
Pediatr 1990; 117: 273-282. c

59. Veness-Meehan KA, Richter S, Davis JM. Pulmonary function testing prior to extubation in infants with respiratory distress syndrome. Pediatr Pulmonol 1990; 9: 2-6.

60. Balsan MJ, Jones JG, Watchko JF, Guthrie RD. Measurements of pulmonary mechanics prior to the elective extubation of neonates. Pediatr Pulmonol 1990; 9: 238- 243.

61. Silos EM, Veber M, Schulman M, Krauss AN, Auld P A. Characteristics associated with successful weaning in ventilator-dependent preterm infants. Am
J Perinatol 1992; 9: 374-377.

62. Fox GF, Alexander J, Marsh MJ, Milner AD. Response to added dead space in ventilated preterm neonates and outcome of trial of extubation. Pediatr Pulmonol 1993; 15: 298-303.

63. Laubscher B, can Melle G, Fawer CL, Sekarski N, Calame A. Hemodynamic changes during high frequency oscillation for respiratory distress syndrome. Arch Dis Child 1996; 74: F172-F176.

64. Keszler M, Modanlou HD, Brudno DS, Clark FI, Cohen RS, Ryan RM, et al. Multicenter controlled clinical trial of high-frequency jet ventilation in preterm infants with uncomplicated respiratory distress syndrome. Pediatrics 1997; 100: 593-599.

65. Carlo WA, Siner B, Chatburn RL, Robertson S, Martin RJ. Early randomized intervention with high frequency jet ventilation in respiratory distress syndrome. J
Pediatr 1990; 117: 765-770.

66. Wiswell TE, Graziani LJ, Kornhauser MS, Cullen J, Merton DA, McKeel, et al. High-frequency jet ventilation in the early management of respiratory distress syndrome is associated with a greater risk for ad- verse outcomes. Pediatrics 1996;'98: 1035- 1043.

67. Pardou A, Vermeylen D, Muller MF, Detemmerman D. High-frequency ventilation and conventional mechanical ventilation in newborn babies with respiratory distress syndrome: A prospective, randomized trial. Intensive Care Med 1993; 19: 406-410.

68. The HlFI Study Group. High-frequency oscillatory ventilation compared with conventional mechanical ventilation in the treatment of respiratory failure in preterm infants. New Engl
J Med 1989; 320: 88-93.

69. Abbasi S, Bhutani VK, Spitzer AT, Fox WW. Pulmonary mechanics in preterm neonates with respiratory failure treated with high-frequency oscillatory ventilation compared with conventional mechanical ventilation. Pediatrics 1991; 87: 487-493.

70. Clark RH, Gerstmann DR, Null DM, deLemos RA. Prospective randomized comparison of high-frequency oscillatory and conventional ventilation in respiratory distress syndrome. Pediatrics 1992; 89: 5-12.

71. Ogawa Y, Miyasaka K, Kawano T, Imura S, Inukai K, Okuyama K, et al. A multicenter randomized trial of high frequency oscillatory ventilation as compared with conventional mechanical ventilation in preterm infants with respiratory failure. Early Hum Dev 1993; 32: 1-10.

72. HiFO Study Group. Randomized study of high-frequency oscillatory ventilation in infants with severe respiratory distress syndrome.
J Pediatr 1993; 122: 609-619.

73. Battisti O, Langhendries JP, Francois A, Clark RH. The Provo multicenter early high-frequency oscillatory ventilation trial: Improved pulmonary and clinical outcome in respiratory distress syndrome. Pediatrics 1996; 98: 1044-1057.

74. Bryan AC, Froese AB. Reflections on the HIFI trial. Pediatrics 1991; 87: 565- 567.

75. Chan V, Greenough A. Determinants of oxygenation during high frequency oscillation. Eur
J Pediatr 1993; 152: 350-353.

76. Gerstmann DR, Fouke JM, Winter DC, Taylor AF, deLemos RA. Proximal, tracheal, and alveolar pressures during high-frequency oscillatory ventilation in a normal rabbit model. Pediatr Res 1990; 28: 367-373.

77. Dimitriou G, Greenough A. Measurement of lung volume and optimal oxygenation during high frequency oscillation. Arch Dis Child 1995; 72: F180-F183.

78. Keszler M, Donn SM, Bucciarelli RL, Alverson DC, Hart M, Lunyong V, et al. Multicenter controlled trial comparing high-frequency jet ventilation in newborn infants with pulmonary interstitial emphysema.
J Pediatr 1991; 119: 85-93.

79. Nelle M, Zilow EP, Linderkamp O. Effects of high-frequency oscillatory ventilation on circulation in neonates with pulmonary interstitial emphysema or RDS. Intens Care Med 1997; 23: 671-676.

80. Kohlhauser C, Popow C, Helbich T, Hermon M, Weninger M, Herold CJ. Successful treatment of severe neonatal emphysema by high-frequency oscillatory ventilation. Pediatr Pulmonol 1995; 19: 52-55.

81. Pappas MD, Sarnaik AP, Meert KL, Hasan RA, Lieh-Lai MW. Idiopathic pulmonary hemorrhage in infancy. Clinical features and management with high frequency ventilation. Chest 1996; 110: 553- 555.

82. Fok TF, Ng PC, Wong W, Lee CH, So KW. High frequency oscillatory ventilation in infants with increased intra-abdominal pressure. Arch Dis Child 1997; 76: F123- F125.

83. Chan V, Greenough A, Gamsu HR. High frequency oscillation for preterm infants with severe respiratory failure. Arch Dis Child 1994; 70: F44-F46.

84. Paraka MS, Clark RH, Yoder BA, Null DM Jr. Predictors of failure of high-frequency oscillatory ventilation in term infants with severe respiratory failure. Pediatrics 1995; 95: 400-404.

85. Ramsden CA, Pillow JJ. High frequency ventilation.
J Pediatr Child Health 1997; 33: 85-87.
 

Home

Past Issue

About IP

About IAP

Feedback

Links

 Author Info.

  Subscription